CVE-2026-66098
Mira Hormone Monitor, Mira Android App Missing authentication for critical function
CVSS Score
6.5
EPSS Score
0.0%
EPSS Percentile
0th
The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.
| CWE | CWE-306 |
| Vendor | quanovate tech inc. (operating as mira / mira care) |
| Product | mira firmware |
| Published | Aug 11, 2026 |
Stay Ahead of the Next One
Get instant alerts for quanovate tech inc. (operating as mira / mira care) mira firmware
Be the first to know when new medium vulnerabilities affecting quanovate tech inc. (operating as mira / mira care) mira firmware are published โ delivered to Slack, Telegram or Discord.
Get Free Alerts โ
Free ยท No credit card ยท 60 sec setup
CVSS v3 Breakdown
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
None
Availability
High
Affected Versions
Quanovate Tech Inc. (operating as Mira / Mira Care) / Mira Firmware
1.7.1.47
Quanovate Tech Inc. (operating as Mira / Mira Care) / Mira Android App
4.5.15.4
References
Credits
Gigi Xiaoqing Liu, Muzzammil Mohammed, Narmina Karimova, and En Mong of Northeastern University SPQR Lab reported this vulnerability to Quanovate Tech.