Flow Neuroscience FL-100 Use of Hard-coded Credentials

CVSS Score
8.1
EPSS Score
0.0%
EPSS Percentile
0th

An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state.

CWE CWE-798
Vendor flow neuroscience
Product fl-100
Published Aug 13, 2026
Stay Ahead of the Next One

Get instant alerts for flow neuroscience fl-100

Be the first to know when new high vulnerabilities affecting flow neuroscience fl-100 are published โ€” delivered to Slack, Telegram or Discord.

Get Free Alerts โ†’ Free ยท No credit card ยท 60 sec setup

CVSS v3 Breakdown

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
High
Availability
High

Affected Versions

Flow Neuroscience / FL-100
0 < July 2026
Halo Neuroscience / FL-100
0 < July 2026

References

NVD โ†— CVE.org โ†— EPSS Data โ†—
cisa.gov: https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-225-01 github.com: https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-225-01.json

Credits

A.C. Buglione reported this vulnerability to CISA.