🔐 CVE Alert

CVE-2025-15653

MEDIUM 6.8

Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation

CVSS Score
6.8
EPSS Score
0.0%
EPSS Percentile
6th

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

CWE CWE-668
Vendor dräger
Product zeus ie
Published Jun 2, 2026
Last Updated Jun 3, 2026
Stay Ahead of the Next One

Get instant alerts for dräger zeus ie

Be the first to know when new medium vulnerabilities affecting dräger zeus ie are published — delivered to Slack, Telegram or Discord.

Get Free Alerts → Free · No credit card · 60 sec setup

CVSS v3 Breakdown

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
Physical
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

Affected Versions

Dräger / Zeus IE
0 < 1.0.5
Dräger / Zeus RS C500
0 < 1.0.9

References

NVD ↗ CVE.org ↗ EPSS Data ↗
static.draeger.com: https://static.draeger.com/security/download/Product-Security-Advisory-25-349-Zeus.pdf vulncheck.com: https://www.vulncheck.com/advisories/dr-ger-zeus-ie-anesthesia-workstation-usb-interface-privilege-escalation