🔐 CVE Alert

CVE-2019-25723

MEDIUM 4.0

Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface

CVSS Score
4.0
EPSS Score
0.0%
EPSS Percentile
14th

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

CWE CWE-1286
Vendor dräger
Product perseus a500
Published Jun 2, 2026
Last Updated Jun 3, 2026
Stay Ahead of the Next One

Get instant alerts for dräger perseus a500

Be the first to know when new medium vulnerabilities affecting dräger perseus a500 are published — delivered to Slack, Telegram or Discord.

Get Free Alerts → Free · No credit card · 60 sec setup

CVSS v3 Breakdown

CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Attack Vector
Network
Attack Complexity
High
Privileges Required
None
User Interaction
None
Scope
Changed
Confidentiality
None
Integrity
None
Availability
Low

Affected Versions

Dräger / Perseus A500
2.00 2.01 2.02

References

NVD ↗ CVE.org ↗ EPSS Data ↗
static.draeger.com: https://static.draeger.com/security/download/2020-06-PSA-20-169-01-Security-Advisory-Perseus-A500-SW20n.pdf vulncheck.com: https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interface