Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.

CWE CWE-400
Vendor dräger
Product infinity m300
Published Jun 2, 2026
Last Updated Jun 3, 2026
Stay Ahead of the Next One

Get instant alerts for dräger infinity m300

Be the first to know when new medium vulnerabilities affecting dräger infinity m300 are published — delivered to Slack, Telegram or Discord.

Get Free Alerts → Free · No credit card · 60 sec setup

CVSS v3 Breakdown

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
None
Availability
High

Affected Versions

Dräger / Infinity M300
0 ≤ VG2.3.1

References

NVD ↗ CVE.org ↗ EPSS Data ↗
static.draeger.com: https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf vulncheck.com: https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service