🔐 CVE Alert

CVE-2019-25719

HIGH 8.6

Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

CVSS Score
8.6
EPSS Score
0.0%
EPSS Percentile
4th

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

CWE CWE-924
Vendor dräger
Product infinity acute care system
Published Jun 2, 2026
Last Updated Jun 3, 2026
Stay Ahead of the Next One

Get instant alerts for dräger infinity acute care system

Be the first to know when new high vulnerabilities affecting dräger infinity acute care system are published — delivered to Slack, Telegram or Discord.

Get Free Alerts → Free · No credit card · 60 sec setup

CVSS v3 Breakdown

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Attack Vector
Network
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
Low
Integrity
Low
Availability
High

Affected Versions

Dräger / Infinity Acute Care System
0 ≤ VG4.1.1 0 ≤ VG4.0.3
Dräger / Standalone Infinity M540 patient monitor
0 ≤ VG4.1.1

References

NVD ↗ CVE.org ↗ EPSS Data ↗
static.draeger.com: https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf vulncheck.com: https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling